The KUSB-3108 module features a variety of analog input channels, including a Cold Junction Compensation (CJC) channel, as well as single-ended/differential analog input channels. The CJC channel provides 10mV/°C with an accuracy of 1°C. The differential analog input channels can be used for B, E, J, K, N, R, S, or T type thermocouples. To measure other low-level analog input signals, such as pressure and flow sensors, you can use these channels to connect up to 16 single-ended/eight differential analog inputs.
Thermocouple break detection is also provided to set the value to full-scale if an open circuit is detected at the input. In addition, the KUSB-3108 provides two analog output channels for high resolution, single-value output operations, 17 digital I/O lines, and two counter/timer channels. The 500V isolation barrier protects your computer and ensures a reliable stream of data. Low gain, Multifunction - High gain, Multifunction
- USB 2.0 and USB 1.1, true plug-and-play compatibility
- 16-bit resolution
- High and low programmable gains
- Input/output speeds up to 50kS/s
- Up to 17 digital lines for monitoring or control
- 500V isolation barrier protects the computer and module from voltage spikes, ESD, and surges
- Built-in/Removable screw terminals
- Runs under Windows® 2000, XP, Vista, 7
|